The 600T860 Xerox-Weston Analogue Multi-meter
I began my working career with Xerox a number of years ago.ย I had the opportunity to go into field service and took advantage of the the great training, equipment and mentoring available at that time.ย That was where I learned the principles of troubleshooting electronic and mechanical devices.ย Part of my first toolkit was the Xerox standard analogue multi-meter 600T860 (Xerox Tool part number).ย A few years into my career they made the transition to a standard digital multi-meter (Simpson 467) and offered the analogue meters for sale to employees.ย I bought mine, and this is that meter.
I have used the 600T860 fairly regularly over the years, particularly when aligning equipment that needed the analogue touch: Either for circuit loading reasons or peaking/nulling functions. I recently reached for it to check some voltages and noticed that the readings where erratic and inconsistent. The batteries and fuses (one externally accessible and one internal) were OK, but I did note some corrosion product (probably copper carbonate) along some of the traces on the backside of the main switch pcb. This post summarizes my troubleshooting and repair of the meter.
Careful visual inspection of the meter revealed the following:
- PCB trace corrosion product near the battery terminals and several other board locations
- Dirty PCB surfaces which were somewhat sticky and appeared to retain dirt and dust
- Crack solder joint on the switch PCB under the input fuse holder
- Tarnished contacts on the main switch
- Tarnished and mechanically slack board interconnect contacts
By way of background, this meter uses three interconnected PCBs: Two for the main function switch, and on for the front panel push buttons. They are interconnected using board to board pins and jacks in a modular arrangement. The board assembly can be removed from the meter body without soldering and the meter movement itself is sealed from the inside of the case.




Note that the meter movement is sealed from the inside of the case.



I re-soldered any joint or connector that looked suspicious, and cleaned the residual flux from the board.
Then I removed as much corrosion product as possible from the PCBs using stainless steel and stiff polymer brushes. I thoroughly cleaned the PCB surfaces with residue free contact cleaner and rinsed all the surfaces with IPA. This was all done while wearing nitrile gloves, of course.
Finally, I cleaned the switch and board interconnect contacts using DexoIT gold which I applied and collected using appropriate small diameter swabs. This method avoided spraying or slopping cleaner onto the switch wafers and PCB.


After reassembling the meter, all the functions worked as expected. So the problem was likely mechanical (cracked joints due to mechanical stress) and electronic (leakage paths caused by corrosion product and dirt accumulation) and contaminated switch and interconnect contacts.
Of course after all these efforts, the meter was out of calibration. Fortunately, there are only five items on the calibration list for this meter:
- Mechanical Zero
- Full Scale Current (25uA)
- Low voltage DC (full scale)
- Low voltage AC (full scale)
- Mid voltage AC (full scale)
These adjustments were all done with the meter assembled and the front panel plate removed. The adjustment potentiometers are accessible via cutouts in the case located under the removable front panel.
I checked the 600T860 against my calibrated Keysight U1253B on all functions except AC current. The repaired Xerox meter was well within specification on all the ranges checked. For example the low DC voltage ranges are specified to have +-2% FS accuracy:

The high voltage DC ranges (150 V to 6000 V) were checked using a 1000 V DC source (Systron Donner M107), and they were within specification. The high voltage AC ranges (150 VAC to 1000 VAC at 400Hz) were checked using a Valhalla 2703 AC Voltage source. All other voltages and currents were checked using a calibrated HP 3245A (Option 2) Universal Source. Resistance ranges were checked using an esi DB877 resistance box. All these instruments are referenced to a calibrated HP 3458A DVM.
I do not have the official Xerox documentation for this meter, but a quick online search led to a matching set of specifications and calibration checks which looked reasonable for a good quality analog meter, so I used those. I now have confidence that I can use the meter in my lab for many years to come.